Performance Verification Procedures

Description
Digitizing Oscilloscope
Signal Generator Precision
DC voltage source
BNC Adapter
50Ω BNC Feed Through
Adapter
WARNING
WARNING
Keysight N2790A Differential Voltage Probe User Guide
The following procedures can be used to test and verify the DC differential
gain accuracy and bandwidth of the N2790A high voltage differential
probe.
Table 7
Critical Specifications
Bandwidth: >100 MHz
1 MΩ/50Ω selectable input
Amplitude accuracy: less than or equal to
0.25%
1 MΩ / 50Ω selectable load
Sine wave greater than or equal to 25 MHz
BNC (f) to Dual Banana (m) Adapter
50Ω precision feed through
These procedures require the application of high voltage to the inputs of
the N2790A probe. Because the adapter used in this procedure has
exposed metal surfaces, only qualified personnel should perform any
testing with voltage levels exceeding 30V rms. All pertinent safety rules
and guidelines for elevated voltage measurements should be followed
and adhered to.
Generator produces hazardous voltages. To avoid risk of shock, do not
touch exposed metal parts after the generator output is enabled.
Required Test Equipment
Recommended Model Part
Number
Keysight MSO9254A
Fluke 9500B or
Fluke 5520A for DC source.
Keysight 1251-2277
HP 11048C
Performance Verification Procedures
Functions
Display probe output
Signal source for DC gain and
bandwidth
Interconnection between probe
and generator
Termination between probe and
calibrator for bandwidth
verification
23