XM multi-axis calibrator
Care and handling
System
• It is recommended to store the XM in the system case when not in use.
• Do not attempt to clean the system with water or other fluids.
• Avoid exposure to oil and coolant.
• Do not direct the air lines at the XM.
• Do not expose the system to impact.
Conduit
• Ensure the conduit is not pinched, crushed or stretched. If damaged, consult
the
Fibre optic safety
section.
• When storing in the case, twist the launch whilst vertical to allow the conduit
to coil.
• If the laser is left in the case during measurement, do not close the lid of the
case to avoid damage to the conduit.
• Fix the conduit using the magnetic cable clamps to ensure the conduit does
not drag, misaligning the launch.
− Manually check the movement of the axis over the full range of the test,
before running a test.
• Never hold the laser or launch by the conduit.
Optics
Cleaning of the optics should be a last resort
To maintain system performance, the XM optics must be kept clean by
following good handling practice:
• Close the receiver and launch shutters when not in use.
• Do not touch the optical surfaces.
• Minimise use in contaminated atmospheres.
Cleaning recommendations
• Only use approved solvents for cleaning the optics: Methylated spirit and
optical grade IPA only (methylated spirit is preferred to IPA).
• Wipe only with non-abrasive lens tissue or lint-free cloth wrapped around a
cotton bud (do not use a cotton bud directly on the optic as this may increase
debris).
• Clean the optics using a gentle action. Never use a scrubbing action as this
might damage the coatings.
Failure to follow these recommendations may lead to damage to the coatings and
glass elements of the optics.