Receiver Alignment - Renishaw XM-60 multi-axis calibrator User Manual

Xm series multi-axis calibrator
Hide thumbs Also See for XM-60 multi-axis calibrator:
Table of Contents

Advertisement

XM multi-axis calibrator

Receiver alignment

• To maintain the system accuracy the launch and the receiver must be
parallel to each other.
• The software will check the parallelism on the completion of the fine axis
alignment.
• If the parallelism tolerance is not met when selecting Capture, the software
will prompt the user to physically adjust the mounting of the receiver. Rotate
the receiver until the red border disappears. Do not adjust the pitch and yaw
adjustment of the launch unit.
Note: It maybe necessary to perform minor translation of the
machine after this step is complete.
Note: Once set-up is complete and before capturing data, it is
recommended to perform an ambient light check. See Capture user
guide for details.