Test
Meter Range
No.
1
4.000V AC
2
4.000V AC,MIN MAX*
3
4.000V AC
4
4.000V AC
5
4.000V AC
6
4.000V AC
7
40.00V AC
8
40.00V AC
9†††
400.0V AC
10
600V AC
11
4.000V DC,MIN MAX*
12
4.000V DC
13
40.00V DC
14
400.0V DC
15
4.000V DC
16
40.00V DC
17
400.0V DC
18
600V DC
19
Continuity
20
Continuity, VCHEK
21
Continuity, VCHEK
22
400.0Ω
23
400.0Ω
24
400.0Ω
25
4.000 kΩ
26
40.00 kΩ
27
400.0 kΩ
28
4.000 MΩ
29
40.00 MΩ
Ω, VCHEK
30
31††
1.000µF
32††
1.000µF
*
MIN MAX tests are for Fluke 12 only.
**
Calibrator 50Ω divider override.
The dc volts annunciator must be on.
Conducting performance tests of the 400Ω, 4 kΩ, 40 kΩ, and 1 µF ranges (tests no. 22, 23, 24,
††
25, 26, 31, and 32) verifies that the discrete and integrated circuitry needed to support the other
capacitance ranges are working within specifications. Therefore, these tests indirectly verify that the
meter will meet specification in the 10 µF, 100 µF, 1000 µF, and 10,000 µF ranges.
†††
In the V-CHEK mode, the UUT uses a low-impedance thermistor (approx. 2.5 kΩ) for circuit
protection and load testing (referred to as lo-Z input circuitry). When using the 5100B or 5700A to drive
the UUT with higher voltages, avoid an overload/current limit condition by gradually stepping the voltage
up (waiting two seconds between each step) from 90.0V, 120.0V, 180.0V and 350.0V at 80 Hz each step.
Table 3-3. Performance Tests for Fluke 11 and 12
Input To
Meter
0V
0V
10 mV, 50 Hz
10 mV, 400 Hz
3.5V, 50 Hz
3.5V, 400 Hz
35V, 50 Hz
35V, 400 Hz
350V, 80 Hz
600V, 400 Hz
0V
0V
+20 mV DC
-200mV DC
+3.5V
-35V DC
+100V DC
-600V DC
+2V DC**
+3.43V DC**
-0.5V DC**
0.0Ω
1.0Ω
350.0Ω
1.0 kΩ
35 kΩ
100 kΩ
2.9 MΩ
35 MΩ
1.96V DC
0.0µF
0.95µF
Slide-Switch
Display
Position
Minimum
Center
0.000
Center
0.000
Center
0.007
Right
0.007
Right
3.430
Center
3.430
Center
34.30
Right
34.30
Right
343.0
Center
586
Center
-0.012
Right
-0.002
Right
00.01
Center
-000.3
Center
+3.466
Center
-35.33
Center
+99.0
Center
-606
Right
1.980
Right
3.397†
Right
-0.507†
Right
0.0
Right
0.8
Right
346.6
Right
0.990
Right
34.67
Right
99.0
Right
2.873
Right
34.44
Right
1.940†
Right
-0.001µF
Right
0.93µF
Maintenance
3
Calibration
Display
Maximum
0.003
0.040
0.013
0.013
3.570
3.570
35.70
35.70
357.0
614
0.012
0.002
00.03
-000.1
+3.534
-34.67
+101.0
-594
2.020
3.463†
-0.493†
0.2
1.2
353.4
1.010
35.33
101.0
2.927
35.56
1.980†
0.001µF
0.970µF
3-9