Fluke 10 Service Manual page 15

Hide thumbs Also See for 10:
Table of Contents

Advertisement

10, 11, 12 Multimeters
Service Manual
Maximum Voltage
Between any Terminal
and Earth Ground
Display
Operating Temperature
Storage Temperature
Temperature Coefficient
Relative Humidity
Battery Type
Battery Life
Shock, Vibration
Size (HxWxL)
Weight
Safety
EMI Regulations
Function
Volts AC
Volts DC
Resistance
Capacitance
Continuity/Diode
Test
* The 4000 mV range can only be entered in the manual range mode. Use the 4000 mV range with
accessories.
** The beeper is guaranteed to come on at <25Ω and turn off at >250Ω. The meter detects opens or shorts
of 250 µs or longer.
1-4
Table 1-1. Specifications
600V ms
3 3/4-digits, 4000 counts, updates 4/sec
-10°C to 50°C
-30°C to 60°C indefinitly (to -40°C for 100 hrs)
0.1 x (specified accuracy)/°C (<18°C or >28°C)
0% to 90% (-10°C to 35°C) 0% to 70% (35°C to 50°C)
9V, NEDA 1604 or IEC 6F22
650 continious hours with alkaline 450 continious hours with carbon-zinc
1 meter shock. Per MIL-T-28800D for a Class 3 Instrument
1.35 in x 2.75 in x 5.55 in (3.46 cm x 7.05 cm x 14.23 cm)
10 oz (286 g)
Designed to Protection Class II requirement of UL1244, ANSI/ISA-S82,
CSA C22.2 No 231, and VDE 0411, and IEC 1010 overvoltage category
III.
Complies with FCC Part 15, Class B, and VDE 0871B.
Range
Resolution
4000 mV*
1 mV
4.000V
0.001V
40.00V
0.01V
400.0V
0.1V
600V
1V
4000 mV
1 mV
4.000V
0.001V
40.00V
0.01V
400.0V
0.1V
600V
1V
400.0Ω
0.1Ω
4.000 kΩ
0.001 kΩ
40.00 kΩ
0.01 kΩ
400.0 kΩ
0.1 kΩ
4.000 MΩ
0.001 MΩ
40.00 MΩ
0.01 MΩ
1.000 µF
0.001 µF
10.00 µF
0.01 µF
100.0 µF
0.1 µF
1000 µF
1 µF
10000 µF
10 µF
2.000V
0.001V
Accuracy (50 to 400 Hz)
Model 10
Models 11/12
±(2.9%+3)
±(1.9%+3)
±(2.9%+3)
±(1.9%+3)
±(2.9%+3)
±(1.9%+3)
±(2.9%+3)
±(1.9%+3)
±(2.9%+3)
±(1.9%+3)
±(1.5%+2)
±(0.9%+2)
±(1.5%+2)
±(0.9%+2)
±(1.5%+1)
±(0.9%+1)
±(1.5%+1)
±(0.9%+1)
±(1.5%+1)
±(0.9%+1)
±(1.5%+2)
±(0.9%+2)
±(1.5%+1)
±(0.9%+1)
±(1.5%+1)
±(0.9%+1)
±(1.5%+1)
±(0.9%+1)
±(1.5%+1)
±(0.9%+1)
±(1.9%+3)
±(1.5%+3)
NA
±(1.9%+2)
NA
±(1.9%+2)
NA
±(1.9%+2)
NA
±(1.9%+2)
NA
±(10%+90) Typical
±(1.5%+2)
±(0.9%+2)

Advertisement

Table of Contents
loading

This manual is also suitable for:

1112