Fluke 10 Service Manual page 31

Hide thumbs Also See for 10:
Table of Contents

Advertisement

10, 11, 12 Multimeters
Service Manual
Test No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
3-8
Table 3-2. Performance Tests for Fluke 10
Meter Range
Input To Meter
4.000V AC
0V
4.000 AC
10 mV, 50 Hz
4.000 AC
10 mV, 400 Hz
4.000 AC
3.5V, 50 Hz
4.000 AC
3.5V, 400 Hz
40.00V AC
35V, 50 Hz
40.00V AC
35V,400 Hz
400.0V AC
350V, 80Hz
600V AC
600V, 400Hz
4.000V DC
0V
40.00V DC
+20 mV DC
400.0V DC
-200 mV DC
4.000V DC
+3.5V
40.00V DC
-35V DC
400.0V DC
+100V DC
600V DC
-600V DC
Continuity
+2V DC
400.0Ω
0.0Ω
400.0Ω
1.0Ω
400.0Ω
350.0Ω
4.000 kΩ
1.0 kΩ
40.00 kΩ
35 kΩ
400.0 kΩ
100 kΩ
4.000 MΩ
2.9 MΩ
40.00 MΩ
35 MΩ
Slide-Switch
Display
Position
Minimum
Center
0.000
Center
0.007
Center
0.007
Center
3.395
Center
3.395
Center
33.95
Center
33.95
Center
339.5
Center
580
Center
0.002
Center
00.01
Center
-000.3
Center
+3.445
Center
-35.54
Center
+98.4
Center
-610
Right
1.968
Right
0.0
Right
0.8
Right
344.5
Right
0.984
Right
34.46
Right
98.4
Right
2.855
Right
34.30
Display
Maximum
0.003
0.013
0.013
3.605
3.605
36.05
36.05
360.5
620
0.002
00.03
-000.1
+3.555
-34.46
+101.6
-590
2.032
0.2
1.2
355.5
1.016
35.54
101.6
2.945
35.70

Advertisement

Table of Contents
loading

This manual is also suitable for:

1112

Table of Contents