LaMotte COD3 Plus Operator's Manual page 9

Colorimeter
Hide thumbs Also See for COD3 Plus:
Table of Contents

Advertisement

4. Press
ENTER
test.
5. Follow the test procedure in the
manual to test the prepared
standard. Insert the blank into
the chamber. Close the lid. Press
to scan the blank. The
ENTER
screen will display Scan Blank
Blank Done for about 1 second
and then return to the Test
Menu.
6. Insert the reacted standard
solution into the chamber. Close
the lid. Press
sample. The screen will display
Scan Sample Sample Done for
about 1 second. The result will
appear on the screen.
7.
The displayed result can now be
standardized. Press
to scroll to Calibrate.
8. Press
ENTER
A reverse font (light background
with dark characters) will appear
to indicate that the reading can
be adjusted.
COD3 Plus Colorimeter 10.18
to select the
to scan the
ENTER
or
to select Calibrate.
007 Ammonia-N LRS
Scan Blank
Scan Sample
12:00:00
001/500
007 Ammonia-N LRS
Scan Blank
Scan Sample
12:00:00
001/500
007 Ammonia-N LRS
0.28 ppm
Scan Blank
Scan Sample
12:00:00
001/500
007 Ammonia-N LRS
0.28 ppm
%T/Abs
Calibrate
12:00:00
001/500
007 Ammonia-N LRS
0.28 ppm
, =Edit, ENTER=Save
+ENTER=Default
12:00:00
001/500
9