How Do I Parallel Channels To Output Higher Current - Keithley 2601A User Manual

2600a series system sourcemeter
Table of Contents

Advertisement

Series 2600A System SourceMeter
Figure A-3
Stacking channels for higher voltage
NOTE Each stacked channel adds approximately 100μA of common mode

How do I parallel channels to output higher current?

The maximum DC output current of a Model 2601A or Model 2602A channel is as follows:
1.01A @ up to 40.4V
3.03A @ up to 6.06V
The maximum DC output current of a Model 2611A/2612A or Model 2635A/2636A channel is as
follows:
101mA @ up to 202V
1.515A @ up to 20.2V
Higher current can be output by connecting channels in parallel.
of two Model 2602As, 2612As, or 2636As connected in parallel. Two Model 2602As can output up
to 4.04A at up to 40.4V and 12.12A at up to 6.06V, while two Model 2612As and Model 2636As
can output up to 404mA at up to 202V and 6.06A at up to 20.2V. The current delivered to the DUT
is the sum of currents output by the four SMU channels. More current can be output by connecting
additional channels in parallel.
2600AS-900-01 Rev. B / September 2008
®
Instruments User's Manual
+
+
+
+
current that is seen by the channels below it. Therefore, in the
example above, Model 2602A-1 Chan A will measure approximately
100μA x 3 = 300μA higher than Model 2602A-2 Chan B. Similarly,
Model 2602A-1 Chan B will measure approximately 100μA x 2 =
200μA higher than Model 2602A-2 Chan B. The most accurate
measurement of actual current flowing through the DUT will be
obtained from Model 2602A-2 Chan B.
Return to
Section Topics
WARNING
Note
Figure A-4
Appendix A: Frequently Asked Questions
shows four channels
A-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2635a2602a2611a2612a2636a

Table of Contents