Installation

WARNING: Do not exceed the electrical rating
printed on the data plate located on the upper left
side of the unit.

Location

Install the unit on a level area free from vibration with minimum
3 inches of clearance on all sides and 4 inches on top. Do not
position the equipment in direct sunlight or near heating
diffusers, radiators, or other sources of heat.
Note: Do not move the unit in fully loaded condition.

Leveling the Unit

The refrigerator must be level in order to provide adequate
condensation drainage as well as proper door alignment and
operation. The refrigerator should be in its final operating
location and set so that it is firmly positioned on the floor.
Level the cabinet front to rear and side-to-side using the
corner leveling screws. The front leveling screws are
accessed by removing the base grill, as described below:
1. Remove the lower grill attachment screws
2. Grasp the grill with both hands
3. Lift the grill approximately 1/2"
4. Pull grill away from the refrigerator
Castor Installing and Pallet
Removal Instructions
CAUTION: Avoid damage to the refrigerant
tubing which may cause a refrigerant leak while
handling, moving, and operating this unit.
Note: Casters available for 12, 25, and 30 cu. ft. units;
leveling feet only on 45, 49, and 72 cu. ft. units
8 |
Installation
Without Forklift
Required Tools: Phillips head screwdriver, 7/16" and 3/8"
wrenches, and a minimum of 2 people.
All the units are shipped with factory installed feet with rolling
casters supplied inside of the unit. The following instructions
demonstrate how to remove the cabinet from its shipping
base (pallet) and install the casters without use of a forklift.
Note: The unit should be unpacked near the desired final
location.
1. Remove the packaging and kick plates from unit.
Remove additional packaging to access the interior of the
cabinet. Be sure to keep one cardboard corner for
subsequent step.
2. Using a screwdriver, remove the screws holding on the
front grill. Lift grill upwards and pull away from unit to
remove. Set aside.
3. Remove bolts (4) holding the unit to the pallet using the
7/16" wrench.
Note: From this step onward, a minimum of 2 people are
required. Use additional people if necessary.
4. With one person on each side, gently slide the unit
towards the side edge of the pallet. Slide the unit so that
General Purpose and Chromatography Refrigerators