2. Warnings and Precautions
2.2
16

Electricity-related Safety Precautions

■ Check the status of the power source, PC and cables prior to operating the
equipment.
■ Ensure that the on/off switch is set to off when the equipment is not in use.
■ Always disconnect the power supply before cleaning the equipment.
■ Always keep electrical cords away from hot appliances or radiators.
■ DO NOT place the PC or peripheral equipment connected to the PC in the
immediate vicinity of the patient.
■ The equipment and PC should be connected to a common protective earth.
■ Never overload the equipment's circuit by sharing it with too many
appliances.
■ Use the same power circuit for the PC and the equipment.
Combining this equipment with other devices
■ Do not connect this equipment to devices which are not part of the system.
■ Do not connect this equipment to a Multiple Portable Socket-Outlet (MPSO)
or extension cord.
Electromagnetic compatibility
■ The PaX-i X-Ray unit complies with IEC standard 60601-1-2.
Medical electrical equipment is subject to special EMC preventive
measures. It must be installed and operated as specified in the Installation
Requirements document.
If high-voltage systems, radio link systems or MRI systems are located
within 5 m of the unit, please observe the specifications stated in the
installation requirements.
Portable and mobile RF communications equipment may interfere with
medical electrical equipment. Therefore, the use of mobile wireless phones
in medical offices or hospital environments must be prohibited.
PaX-i (PCH-2500) User Manual