Recommended X-Ray Exposure Table - Vatech PAX-I User Manual

Hide thumbs Also See for PAX-I:
Table of Contents

Advertisement

Appendices
1.
120

Recommended X-Ray Exposure Table

A. PANO mode
Scan Time Mode: UHD / HD
Gender / Figure
Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
Man
Woman
Child
Scan Time Mode: Normal
Gender / Figure
Hard (kVp / mA) Normal (kVp / mA) Soft (kVp / mA)
Man
Woman
Child
B. CEPH mode
Lateral
Figure
Hard (kVp / mA)
Gender
Sensor
One-shot
Type
Man
90 / 10
Woman
90 / 10
Child
86 / 10
PA / SMV / Waters View
Figure
Hard (kVp / mA)
Gender
Sensor
One-shot
Type
Man
90 / 10
Woman
90 / 10
Child
86 / 10
74 / 10
73 / 10
73 / 10
72 / 10
68 / 8
67 / 8
74 / 8
73 / 8
73 / 8
72 / 8
68 / 8
67 / 8
Normal (kVp / mA)
Scan
One-shot
Scan
86 / 10
90 / 10
85 / 10
85 / 10
90 / 10
84 / 10
81 / 10
85 / 10
80 / 10
Normal (kVp / mA)
Scan
One-shot
Scan
88 / 10
90 / 10
87 / 10
87 / 10
90 / 10
86 / 10
83 / 10
85 / 10
82 / 10
72 / 10
71 / 10
66 / 8
72 / 8
71 / 8
66 / 8
Soft (kVp / mA)
One-shot
Scan
89 / 10
84 / 10
89 / 10
83 / 10
85 / 10
80 / 9
Soft (kVp / mA)
One-shot
Scan
89 / 10
86 / 10
89 / 10
85 / 10
84 / 10
81 / 10