Multi-Signal Module; Testing; Digital Display Module - BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions

Medical electronic gas alarm
Table of Contents

Advertisement

Operation
Monitoring Mode

Multi-Signal Module:

Multi-Signal modules monitor up to five
dry-contact switches.
• If the signal is normal (switch contacts
closed), green indicator for signal is
illuminated.
• If an alarm occurs (switch contacts
open), audible alarm activates and red
indicator for signal flashes.
• When front panel alarm MUTE
button is pressed, audible alarm is
silenced. Alarm indicator remains
illuminated, but does not flash.
• Visual indicator will remain illuminated
as long as alarm condition remains.
• If audible alarm repeat time has been
set, audible alarm will reactivate, and
visual indicator will again flash after
specified time.
SIGNAL 2
SIGNAL 3
SIGNAL 4
SIGNAL 5

Testing

Periodic testing of alarm system is
recommended. Digital display modules
and multi-signal alarm modules have built-
in self-test modes. Front panel TEST
button will initiate self-test mode on all
alarm panel modules at same time.
SIGNAL 1
Front panel
TEST button

Digital Display Module:

Press and hold the front panel TEST but-
ton for five seconds to initiate the following
sequence:
Audible alarm sounds briefly as T T E E S S T T is
displayed.
All LED display segments and LOW,
NORMAL, and HIGH indicators flash for
five seconds (firmware up to revision
1.0.3).
LED display segments and LOW,
NORMAL, and HIGH indicators illuminate
in sequence (firmware revision 1.0.4 and
later).
Firmware revision (e.g. 1 1 . . 0 0 . . 6 6 ) is displayed
for two seconds.
A A D D D D R R is briefly displayed and then current
MEGA network module address is
displayed for two seconds. (Factory default
address is 0 0 ).
Current units of measure setting (e.g. P P S S I I )
is displayed for two seconds.
L L O O is briefly displayed, LOW alarm indicator
is illuminated, and current low alarm set
point is displayed for two seconds.
K K , , is briefly displayed, HIGH alarm
indicator is illuminated, and current high
alarm set point is displayed for two
seconds.
Module returns to monitoring mode.
Part No. 6-847684-00 Rev. D01
41