Multi-Signal Module - BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions

Medical electronic gas alarm
Table of Contents

Advertisement

Operation
Set-Up Procedure

Multi-Signal Module

Changing signal options:
1. Loosen two screws on front of alarm
panel. Hinge open front cover.
2. Locate SETUP button on circuit board
of multi-signal module to be adjusted.
3. Press and release SETUP button.
Audible alarm will chirp and both
green and red indicators for top
location (Signal 1) will illuminate.
4. Using either
or
or disable Signal 1 location. When red
indicator is illuminated, location is
ENABLED. Turning off red indicator
disables location. Signal locations can
not be disabled if wired to a closed-
contact circuit.
5. Press and release TEST button on
front of alarm panel. Both green and
red indicators for second location
(Signal 2) will be illuminated.
6. Using either
or
or disable Signal 2 location. When red
LED is illuminated, location is enabled.
Turning off red LED disables location.
7. Repeat steps 5 and 6 above for
remaining signal locations (Signal 3, 4,
and 5).
8. Press and release TEST button to
save changes. LEDs will sweep from
top to bottom to indicate end of set-up
sequence has been reached. Multi-
signal module will return to monitoring
mode.
38
Part No. 6-847684-00 Rev. D01
buttons, enable
buttons, enable
Multi-Signal Alarm Module Circuit Board
SETUP
button
Red LED
SIGNAL 1
SIGNAL 2
SIGNAL 3
SIGNAL 4
SIGNAL 5
TEST
button
NOTE:
Audible alarm will chirp with each button push.
If module is idle for more than one minute during
set-up procedure, module will chirp three times,
return to monitor mode and will default to previ-
ous settings. No changes to settings will be
saved.
The set-up must be completed in order for
changes to be saved.
Green LED
Up and down
buttons