BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions page 37

Medical electronic gas alarm
Table of Contents

Advertisement

Operation
Set-Up Procedure
Digital Display Module
R R E E P P T T
(audible alarm repeat)
Used to set time interval for
reactivation of audible alarm after
alarm MUTE
pressed. User choices are O O F F F F , 1 1 , 5 5 ,
1 1 5 5 , 3 3 0 0 (minutes), 1 1 H H R R , 8 8 H H R R , 1 1 2 2 H H R R or
2 2 4 4 H H R R (hour). Factory setting is O O F F F F .
When set to O O F F F F , audible alarm will
not reactivate after an active alarm
has been silenced. Any new alarm will
reactivate audible alarm.
I I N N T T S S
(display intensity)
Used to adjust digital module's display
brightness. Factory setting is
maximum intensity.
NOTE:
If module is idle for more than one minute during
set-up procedure, module will chirp three times,
return to monitor mode and will default to previous
settings. No changes to settings will be saved.
Set-up must be completed and module must dis-
D D O O N N E E
play
in order for changes to be saved.
Changing Communication Parameters -
(S S E E T T 2 2 ):
1. Press and HOLD SETUP button for
TWO seconds until S S E E T T 2 2 is displayed.
2. Press and release TEST button on
front of alarm panel. A A D D D D R R will be
displayed for one second and then
current network address. Select
desired address by pressing
buttons. Each module must have a
unique address.
3. Press and release TEST button.
B B A A U U D D will be displayed for one second
and then current baud rate. Select
desired baud rate by pressing
buttons.
4. Press and release TEST button. N N D D F F
will be displayed for one second and
(Advanced)
button has been
or
or
then current network data format.
Select desired data format by
pressing
or
5. Press and release TEST button. P P D D F F
will be displayed for one second and
then current pressure data format.
Select desired data format by
pressing
or
6. Press and release TEST button to
save changes. D D O O N N E E will be
displayed for two seconds. Digital
display module will return to
monitoring mode.
Changing Miscellaneous Parameters -
(S S E E T T 3 3 ):
1. Press and HOLD SETUP button for
FOUR seconds until S S E E T T 3 3 is
displayed.
2. Press and release TEST button on
front of alarm panel. S S E E R R U U will be
displayed for one second and then
current gas service identification.
Select desired gas service by
pressing
or
3. Press and release TEST button.
O O F F F F S S will be displayed for one second
and then current display offset. Select
offset by pressing
4. Press and release TEST button.
R R E E P P T T will be displayed for one second
and then current audible alarm repeat
time. Select desired repeat time by
pressing
or
5. Press and release TEST button. I I N N T T S S
will be displayed. Select desired
display brightness by pressing
buttons. Each push will change
brightness of display. Audible alarm
will chirp one time with each push and
two times when limit reached.
6. Press and release TEST button to
save changes. D D O O N N E E will be
displayed for two seconds. Digital
display module will return to
monitoring mode.
buttons.
buttons.
buttons.
or
buttons.
buttons.
or
Part No. 6-847684-00 Rev. D01
37