BeaconMedaes MEGA 2 Installation, Operation And Maintenance Instructions page 16

Medical electronic gas alarm
Table of Contents

Advertisement

Wiring Schematic 3: Multi-Signal Module
Dry normally-closed
contacts at source
equipment or line
pressure switches
+
COM
SIGNAL 1
+
COM
SIGNAL 2
+
COM
SIGNAL 3
COM
+
SIGNAL 4
+
COM
SIGNAL 5
Multi-Signal Module front panel
16
Part No. 6-847684-00 Rev. D01
Pin 10
Multi-Signal Module
at FIRST master alarm
panel
NOTE:
Installer assigns source signals to desired front panel locations
and installs provided labels (Refer to Figure 23, page 31).
NOTE:
Field wiring cable shields must be grounded at
only one end, inside alarm panel back box. Refer
to page 31 for details.
Pin 10
Pin 1
ribbon to
adjacent
module
at SECOND master alarm panel
SIGNAL 1 - pin 1 (+), pin 2 (COM)
SIGNAL 2 - pin 3 (+), pin 4 (COM)
SIGNAL 3 - pin 5 (+), pin 6 (COM)
SIGNAL 4 - pin 7 (+), pin 8 (COM)
SIGNAL 5 - pin 9 (+), pin 10 (COM)
Pin 1
Multi-Signal Module
ribbon to
adjacent
module