Wiring Schematic 6: General Fault / Auxiliary Relays
Auxiliary Relay
N/O (pin 6)
Common (pin 7)
N/C (pin 8)
Annunciator Module
Wiring Schematic 7: Digital Display Module High / Low Relays
High Alarm Relay
Common (pin 9)
N/C (pin 10)
Digital Display Module
General Fault Relay
N/C (pin 5)
Common (pin 4)
N/O (pin 3)
Low Alarm Relay
N/C (pin 8)
Common (pin 7)
NOTE:
General Fault Relay activates when ANY
alarm on panel is in progress. Pressing
MUTE
NOT reset relay. Relay deactivates when
alarm condition is corrected.
Auxiliary Relay (optional) activates when
ANY audible alarm on panel is in progress.
ribbon to
Pressing MUTE
adjacent
WILL deactivate relay until audible alarm is
module
again reactivated.
Relay contact ratings are 2A @ 30 VDC/0.5
A @ 125 VAC.
NOTE:
High Alarm Relay activates when high line
pressure alarm on digital display module is
in progress. When relay activates normally
closed contacts open. Pressing MUTE
on Annunciator Module DOES NOT deacti-
vate relay. Relay deactivates when high line
pressure alarm condition is corrected.
ribbon to
Low Alarm Relay activates when low line
adjacent
pressure alarm on digital display module is
module
in progress. When relay activates normally
closed contacts open. Pressing MUTE
on Annunciator Module DOES NOT deacti-
vate relay. Relay deactivates when low line
pressure alarm condition is corrected.
Relay contact ratings are 2A @ 30 VDC/0.5
A @ 125 VAC.
on Annunciator Module DOES
on Annunciator Module
Part No. 6-847684-00 Rev. D01
19