Weekly
Testing
Transmitter
Cleaning
and Care
Safe Place Transmitters User Guide
0510-1121-G
The following testing is required weekly for
transmitters in use on patients. All steps are
mandatory.
1. Test the operation of transmitters using
the transmitter tester.
2. Visually inspect transmitters for damage
or loose parts.
3. Verify that the warranty expiration date
stamped on the transmitter has not
expired. If the warranty period has
expired, discard and replace the
transmitter immediately.
4. Your facility must keep records of test
and transmitter inspection.
NOTE: Never take a patient to a door to test
their transmitter.
This section provides detailed information
about cleaning and caring for a transmitter.
To clean a transmitter after use:
1. Remove the banding material from the
transmitter and dispose of it.
CAUTION: Use of non-approved cleaners
on transmitters can cause permanent
damage to the transmitter and the piercing
contacts and will void any and all
warranties.