TESTING AND CARE
Transmitter
Testing
Visual
Inspection
Safe Place Transmitters User Guide
0510-1121-G
You must test all transmitters prior to use to
verify proper operation. This includes every
time that the band is replaced.
All steps are mandatory.
WARNING: Do not use the transmitter tester
on a transmitter with an expiration date that
has expired. Results may not be accurate.
Do not use a transmitter beyond its
expiration date.
CAUTION: Failure to test the transmitters
before use can result in system failure
and/or an abduction or elopement. In
addition, failure to test transmitters voids the
RF Technologies Product Warranty
1. Verify that the warranty expiration date
that is stamped on the transmitter is not
expired.
2. Visually inspect the transmitter for
damage or loose parts.
3. Attach a band to the transmitter (See
Banding).
Testing and Care
Page 23 of 36