5.1 Calibration

The probe adapter modules carry the calibration data for the instrument. If the unit is shipped with a
probe, the module is calibrated at the factory prior to shipment; otherwise, the module is shipped
uncalibrated. The instrument itself does not require calibration.
If required, calibration data stored in the probe adapter may be changed using the optional
Communicator, P/N 190-1A. Refer to the instruction manual supplied with the Communicator for further
instructions and specific procedures for entering calibration data.
Avoid indiscriminately mating probes with an
adapter
module;
otherwise
instrument or probe or loss of stored calibration
factors,
parameters,
information
may
recommends the use of heat shrink tubing to
permanently connect the probe and adapter.