In the event that the internal and external
EEPROMS were not properly read, an error
message(s) will be displayed (i.e., "INT EEPROM
FAIL" and/or "EXT EEPROM FAIL"). Contact Fluke
Biomedical at 440.248.9300 for further instructions.
d.
The instrument will then begin to operate in the selected mode.
The instrument is factory set with all buttons
operational and the CPM and mR/H modes
selected. (The Log Mode is inactive.)
6. Test each button of the instrument, as follows:
a. Press each LIGHT button to cause the display backlight to turn on. (The backlight may not
show up in a bright environment.)
b. Press the MODE button to toggle the available display modes that were enabled during
calibration. The current mode will be displayed briefly on the 16-character alphanumeric
display. An example:
- COUNTS/MIN MODE
- COUNTS/SEC MODE
- mR/H MODE
If the probe has not been initialized and calibrated
(either at the factory or via the optional infrared
communicator), the CPM and CPS modes are the
only modes available. In addition, the mCi/sqcm,
Bq/sqcm, and dpm modes require an additional
special calibration performed with the probe. Refer
to section 5 - Calibration for further details.
c.
Press the LOG button to log displayed data. If the Log Mode has been activated, the following
should appear in the alphanumeric display:
LOGGING DATA
If the Log Mode has not been activated, the following message will be displayed:
NOT IN LOG MODE
The Log Mode can only be activated using the optional Communicator, Model 190-1A. Refer to the
Instruction Manual supplied with the Communicator for details.
d.
Press the RATE/INTEG button to toggle the alphanumeric display between the Rate Mode
and the Integrate Mode.
e.
Press the RESP TIME button to toggle the available response time selections. The new
selection is displayed in the alphanumeric display as follows:
RESPONSE IS XX S
where XX is the currently active first scale response time.