Specification
Precision
Stray light
Calibration options
Verification options
Verification (RFID or
®
Link2SC
)
Certifications
Warranty

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.

Use of hazard information

Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
4 English
Details
< 40 NTU: 0.002 NTU or 1% (the larger value); > 40 NTU: 3.5% based on
Formazin primary standard at 25 °C (77 °F)
< 0.01 NTU
®
StablCal
: 1-point calibration (20 NTU) for 0 to 40 NTU measurement range; 2-
point calibration (20 and 600 NTU) for 0 to 700 NTU (full) measurement range
Formazin: 2-point calibration (20 NTU and dilution water) for 0 to 40 NTU
measurement range; 3-point calibration (20 NTU, 600 NTU and dilution water) for
0 to 700 NTU (full) measurement range
Degrees: 3-point calibration (20 and 100 mg/L and dilution water) for 0 to 100 mg/L
(full) measurement range
SDVB: 3-point calibration (20 NTU, 600 NTU and dilution water) for 0 to 700 NTU
(full) measurement range
Custom: 2- to 6-point custom calibration for a measurement range of 0 NTU to the
highest calibration point.
Glass verification rod (secondary turbidity standard) < 0.1 NTU, StablCal or
Formazin (0.1 to 40 NTU)
Process and laboratory measurements are compared with RFID or Link2SC for
verification of the measurement value.
CE compliant; US FDA accession number: 1420493-xxx. This product complies
with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No.
50. Australian RCM.
1 year (EU: 2 years)