GE Masoneilan 12420 Instruction Manual & Safety Manual page 37

Masoneilan 12400 series. level transmitter/controller
Table of Contents

Advertisement

f.
H
[SPAN]
IGH LEVEL REFERENCE
f1. Liquid application:
Press the flexure (59) against the adjusting screw shoulder (114) which corresponds to the high level (on displacer
side). While maintaining contact and using a 3 mm hex wrench, turn the adjusting screw until LCD displays the value
of process specific gravity [SG CALIB].
f2. Interface application:
Press the flexure (59) against the adjusting screw shoulder (114) which corresponds to the highest specific gravity
[HSG CALIB]. While maintaining contact and using a 3 mm hex wrench, turn the adjusting screw until LCD displays
this specific gravity.
g. Slowly move the flexure (59) from one shoulder (114) to the other (slowly to avoid displacer oscillations) and verify the value
displayed. Correct adjustments if necessary.
h. Reinstall all plugs (107), (190), and (115).
® 2014 General Electric Company. All rights reserved.
Masoneilan 12400 Series Transmitter/Controller Instruction Manual | 35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Masoneilan 12410Masoneilan 12430

Table of Contents