Functional Checks And Tests; External Sight Control - Welch Allyn CP300 Service Handbook

12 channel ecg recorder
Table of Contents

Advertisement

Service handbook CP300
6.3

Functional Checks and Tests

6.3.1

External Sight Control

Required equipment
• None
6-monthly
Check the following:
• Mechanical condition of the device:
– no cracks or chips in the casing.
– mains, patient and all other cable assemblies are in good condition with no
crushing, chafing or cuts, etc.). All plugs and sockets are straight and in good
condition.
– no soiling which could hamper the safety of the device.
12-monthly
Check the following:
• Mechanical condition of the device as detailed above.
• Voltage selector is set correctly.
• Correct fuse rating according table
• The following safety labels are on the device and are readable:
– Back Panel, type designation and fuse rating label.
– Side Panel (patient connector), CF label and attention symbol.
Functional Checks
Functional Checks and Tests
(see page 68).