Changing The Mains Fuse; Fuse Types; Changing A Fuse - Welch Allyn CP300 Service Handbook

12 channel ecg recorder
Table of Contents

Advertisement

10
Maintenance
10.4

Changing the Mains Fuse

Page 68
10.4
Changing the Mains Fuse
 Before the fuse and mains voltage are changed, the device must be disconnected
from the mains and the mains plug removed for the wall socket.
 The fuse may only be replaced by the fuse type the table below.
10.4.1

Fuse Types

Voltage range
100 - 240 VAC
10.4.2

Changing a Fuse

1.
Disconnect the device from the mains and remove the mains plug.
2.
Loosen the fuse inset by using pushing up the release catch and remove the fuse
inset.
3.
Replace existing fuses with the same type (see table
inset and ensure it clicks in place.
Number
2
250 V / 630 mA (T = slow blow)
Fuse type
above).
Re-insert the fuse