Advertisement

AMI Oxytrace QED
Program List and Explanations
4.2 Relay Contacts
4.3 Logger
4.3.1
Interval
1 s
Time
25 min
4.3.2

5 Installation

5.1 Sensors
5.1.1
5.1.1.1
5.1.1.2
5.1.2
5.1.2.1
74
See
Relay Contacts, p. 30
The instrument is equipped with an internal logger. The data can be
copied to the USB stick installed in the transmitter.
The logger can save approx. 1500 data records. The Records con-
sists of: Date, time, alarms, measuring values, raw values, case
temperature, flow.
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Range: 1 Second to 1 hour
5 s
1 min
2 h
25 h
Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
Miscellaneous
Flow: If a flow cell without flow measurement (e.g. B-Flow) is used,
choose none. With flow measurement select Q-Flow
O2 Offset: Manual, small correction of the offset.
Range -5 to 5 ppb
Quality Assurance: Switch the Quality Assurance on or off.
Level: Select quality level:
 Level 0: Off
Quality assurance procedure switched off. Any additional QA
menus are hidden.
 Level 1: Trend
 Level 2: Standard
 Level 3: Crucial
 Level 4: User
Edit user specific limits in menu 5.1.2.2
5 min
10 min
5 d
10 d
30 min
1 h
31 d
62 d
A-96.250.531 / 300617