AMI Oxytrace QED
Maintenance
Quality Level
0: Off
1: Trend
2: Standard
3: Crucial
4: User
a) sample temperature must have 25°C +/- 5°C.
Procedure
6.8.1
52
max. deviation
temperature [°C]
Off
0.5 °C
0.4 °C
0.3 °C
0 - 2°C
The standard workflow contains following procedures:
1 Activate SWAN Quality assurance procedure
2 Pre-test
3 Connect instruments
4 Carry out comparison measurement
5 Completion of the measurement
NOTICE: The procedure should only be carried out through
qualified personnel.
Materials / Inspection equipment:
Reference instrument: AMI INSPECTOR Oxygen
Two tubes made of PA
Activate SWAN Quality assurance procedure
Enable quality assurance procedure at each instrument by select-
ing the quality level in menu 5.1.4.
The corresponding submenus are then activated.
NOTICE: The activation is necessary the first time only.
max. deviation
a)
result [%]
Off
10 %
5 %
5 %
0 - 20%
min. inspection
interval
Off
annual
quarterly
monthly
annual, quarterly,
monthly
A-96.250.531 / 300617