AMI Oxytrace QED
Installation
3.8.
3.8.1
3.9.
A-96.250.531 / 300617
Signal Outputs
Signal output 1 and 2 (current outputs)
NOTICE: Max. burden 510 Ω.
If signals are sent to two different receivers, use signal isolator
(loop isolator).
Signal output 1:
Signal output 2:
Programming see menu
Interface Options
The slot for interfaces can be used to expand the functionality of
the AMI instrument with either:
an additional signal output
a Profbus or Modbus connection
a HART connection
an USB Interface
Terminals 14 (+) and 13 (-)
Terminals 15 (+) and 13 (-)
5.2 Signal Outputs, p. 75
A
A
B
C
D
B
C
D
AMI Transmitter
Slot for interfaces
Frontend PCB
Screw terminals
33