Signal Outputs; Signal Output 1 And 2 (Current Outputs); Interface Options - Swann AMI Oxytrace Operator's Manual

Hide thumbs Also See for AMI Oxytrace:
Table of Contents

Advertisement

AMI Oxytrace QED
Installation
3.8.
3.8.1
3.9.
A-96.250.531 / 300617

Signal Outputs

Signal output 1 and 2 (current outputs)

NOTICE: Max. burden 510 Ω.
If signals are sent to two different receivers, use signal isolator
(loop isolator).
Signal output 1:
Signal output 2:
Programming see menu

Interface Options

The slot for interfaces can be used to expand the functionality of
the AMI instrument with either:
 an additional signal output
 a Profbus or Modbus connection
 a HART connection
 an USB Interface
Terminals 14 (+) and 13 (-)
Terminals 15 (+) and 13 (-)
5.2 Signal Outputs, p. 75
A
A
B
C
D
B
C
D
AMI Transmitter
Slot for interfaces
Frontend PCB
Screw terminals
33