Technical Data - Swann AMI Oxytrace Operator's Manual

Hide thumbs Also See for AMI Oxytrace:
Table of Contents

Advertisement

AMI Oxytrace QED
Product Description
2.2.
Power Supply
Electronics
housing
Sample
requirements
Flow cell and
connection
On-Site
requirements
Accuracy
Reproducibility
Measuring
Range
Temperature
Sample Flow
Measurement
A-96.250.531 / 300617

Technical Data

Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
Humidity:
Display:
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure
pH:
Suspended solids:
Flow cell made of acrylic glass with built-in flow adjustment valve
and digital sample flow meter
Sample inlet:
Sample outlet:
Sample inlet:
Sample outlet:
± 1.5% of measured value or ± 0.2 ppb
±1% of measured value or ± 0.15 ppb
up to 60°C
Resolution: 0.1°C
With Swan sample flow meter and alarm in case of insufficient
sample flow.
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
-10 to +50 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
8–25 l/h
15–45 °C
0.2 to 1 bar
pressure free
not lower than pH 4
less than 10 ppm
1/4" Swagelok tube adapter
for flexible tube 15x20 mm
Swagelok connection for stainless steel
tube 1/4".
15x20 mm (1/2") hose nozzle which
must end in a pressure free waste of suf-
ficient capacity.
13