Overcompensated
Figure 3-8 Displayed Waveforms of the Probe Compensation
3.
Repeat the steps mentioned if needed.

How to Set the Probe Attenuation Coefficient

The probe has several attenuation coefficients, which will influence the vertical scale
factor of the oscilloscope.
To change or check the probe attenuation coefficient in the menu of oscilloscope:
(1) Press the function menu button of the used channels (CH1 MENU or CH2
MENU).
(2) Press H3 button to display the Probe menu; select the proper value corresponding
to the probe.
This setting will be valid all the time before it is changed again.
Caution:
The default attenuation coefficient of the probe on the instrument is preset to
10X.
Make sure that the set value of the attenuation switch in the probe is the
same as the menu selection of the probe attenuation coefficient in the
oscilloscope.
The set values of the probe switch are 1X and 10X (see Figure 3-10).
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
3.Quick Start
Compensated correctly
Figure 3-9 Adjust Probe
TestEquipmentDepot.com
12
Under compensated