Olympus DELTA 50 Quick Start Manual page 16

Delta family handheld xrf analyzer
Table of Contents

Advertisement

103076-01EN [U8998318], Rev. C, June 2014
Cal Check button
7.
Tap the Start Test button.
The Cal Check takes approximately 15 seconds. After a successful Cal Check, the analyzer is ready for use.
8.
Position the measurement window over the test sample.
9.
Pull the trigger or tap the Start Test button (
Results are displayed immediately upon test completion.
10. Tap the Spectrum button (
11. At the end of the session, export the results to a PC using the data port and USB cable.
12. When testing and exporting are complete, turn off the analyzer using the I/O switch, or insert it into the docking
station.
6.2
About Calibration Check (Cal Check)
DELTA analyzers use advanced techniques for Cal Checking.
The docking station (with its special 316 s/s test cup) provides an automatic Cal Check. The analyzer must be ON
when inserted.
When in the field (and away from the docking station) you can use the 316 s/s standardization coupon to perform a
Cal Check. This check only takes 15 seconds.
16 Operation
Figure 6-2 Cal Check Required message
).
) to view special results (see Figure 6-3 on page 16).
Figure 6-3 Results