MSA ALTAIR 5X Operating Manual page 8

Msha-multigas detector
Hide thumbs Also See for ALTAIR 5X:
Table of Contents

Advertisement

-
®
Use of the GALAXY
calibrating MSHA-approved ALTAIR 5X devices.
-
Use only calibration gas that is 2.5% Methane with an accuracy of +5% when calibrating MSHA-
approved ALTAIR 5X devices.
-
®
GALAXY
GX2 Automated Test Setup Mode must be set to
Cal on Fail
when calibrating MSHA-approved ALTAIR 5X devices.
-
For 30 CFR Part 75 determinations, the GALAXY
must be set to
Calibration Only
detected with an accuracy of +0.5 % when calibrating MSHA-approved ALTAIR 4X devices.
Observe Proper Battery Maintenance
Use only battery chargers made available by MSA for use with this device; other chargers may damage
the battery pack and the device. Dispose of in accordance with local health and safety regulations.
Be Aware of Environmental Conditions
A number of environmental factors may affect the sensor readings, including changes in pressure,
humidity and temperature. Pressure and humidity changes also affect the amount of oxygen actually
present in the atmosphere.
Be Aware of the Procedures for Handling Electrostatically Sensitive Electronics
The device contains electrostatically sensitive components. Do not open or perform maintenance on
the device without using appropriate electrostatic discharge (ESD) protection. The warranty does not
cover damage caused by electrostatic discharges.
NOTE:
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protec-
tion against harmful interference when the equipment is operated in a commercial environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his own
expense.
This is a class A product in accordance with CISPR 22.
In a domestic environment, this product may cause radio interference, in which case the user may
be required to take adequate measures.
This Class A digital apparatus complies with Canadian ICES-003.
Be Aware of the Product Regulations
Follow all relevant national regulations applicable in the country of use.
Be Aware of the Warranty Regulations
The warranties made by Mine Safety Appliances Company with respect to the product are voided if the
product is not used and maintained in accordance with the instructions in this manual. Please protect
yourself and others by following them. We encourage our customers to write or call regarding this
equipment prior to use or for any additional information relative to use or service.
GX2 Automated Test System is an alternate MSHA-approved method for
or
Calibration Check/Cal on Fail
WARNING
ALTAIR 5X MSHA
Calibration Only
®
GX2 Automated Test System Test Setup Mode
so that 19.5 % oxygen can be
Safety Regulations
or
Calibration Check/
US
8