MSA ALTAIR 5X Operating Manual page 41

Msha-multigas detector
Hide thumbs Also See for ALTAIR 5X:
Table of Contents

Advertisement

(1)
Verify that the device is turned OFF.
(2)
Remove the battery pack.
(3)
Remove the two remaining case screws, and remove the case front.
(4)
Gently remove the sensor to be replaced.
(5)
Carefully align the new sensor contact pins with the sockets on the printed circuit board.
(6)
Press the new sensor into place.
(7)
Note the position restrictions in the table above.
Adapter (P/N 10110183) is required for XCell usage in Position #3.
Ensure that a sensor plug is properly installed in any position that does not have a sensor.
The plug for XCell positions is P/N 10105650.
The Series 20 plug is P/N 10088192.
SENSOR
XCell combustible sensor
XCell O
sensor
2
XCell CO-H
S Two-tox sensor
2
XCell SO
, Cl
, NH
2
2
XCell CO-HC
Series 20 sensor
(8)
Re-install the sensor gasket in the case front.
(9)
Re-install the front case, screws, and battery pack.
If a change in XCell Sensor configuration is detected during the device turn-ON process:
-
The "ACCEPT?" prompt appears on the display
-
The  button accepts the sensor configuration
-
The  button rejects the sensor configuration; the device is not operational.
When an XCell sensor is replaced, the device automatically enables the sensor after the change has
been accepted. If a Series 20 sensor is replaced, it must be manually enabled (see section 3.4.3).
If the oxygen sensor was replaced, see section 3.2.3 regarding the oxygen reading display.
(10)
After installation of new sensors, allow them to stabilize at least 30 minutes before calibration.
(11)
Calibrate device before use.
Calibration is required after a sensor is installed; otherwise, the device will not perform as expected
and persons relying on this product for their safety could sustain serious personal injury or death.
OPERATIONAL ONLY IN POSITION
#1
#2
#2 or #4
#3 or #4
3
#4
#3
WARNING
ALTAIR 5X MSHA
Maintenance
US
41