Sample Measurement Shifts Report - Fluke 5790A Operator's Manual

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

5790A
Operators Manual
3. When you print a report, the display appears as follows, allowing your to abort
The printout begins if the interface parameters are set up correctly and the cable wiring is
correct for the peripheral device. If the input buffer of the peripheral device fills and the
device sends the 5790A a stall command, the 5790A stops sending data and starts a timer.
If the timeout period expires, printing is aborted. Press any key to clear the abort
message.
Line Feed characters are embedded in the data stream so that the output is paginated for a
66-line page.
7-5.

Sample Measurement Shifts Report

7-6
printing at any time:
JOHN FLUKE MFG. CO., INC. 5790 CALIBRATION REPORT
Calibration Shifts -- STORED vs. OLD Constants
Printed on 12/17/91 at 11:39:00
Serial Number:
5420001
Report String:
5790A
Software Versions:
Main: 1.2
------------------------------------------------------------------------
CAL TYPE
DATE
Zero
Main Periodic :
Main Service
:
Wideband
:
------------------------------------------------------------------------
NOTE: A positive shift means it takes a larger voltage to produce the same
reading with the new constants than it did with the old constants. An ***
indicates a point calibrated directly during periodic calibration. The I2 vs
I1 shift pertains to measuring the same voltage at INPUT1 and INPUT2. A
positive shift means the new INPUT2 measurement is smaller for the same
INPUT1 measurement.
Inguard: 1.2
CALIBRATION DATES AND TEMPERATURES
ACTIVE DATES AND TEMPERATURE
12/14/91 17:25:47 @ 22.4 degC
12/14/91 18:47:46 @ 22.3 degC
12/14/91 18:47:46 @ 22.3 degC
12/14/91 17:09:17 @ 22.4 degC
MEASUREMENT SHIFTS
2.2 mV RANGE
add55f.eps
add57f.eps
STORED DATE OLD
12/14/91 9/10/91
12/14/91 9/10/91
12/14/91 9/10/91
12/14/91 9/10/91