Placement And Rack Mounting; Accessing The Fuse - Fluke 5790A Operator's Manual

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

The air entering the instrument must be room temperature. Make sure that exhaust
from another instrument is not directed into the fan inlet.
Clean the air filter every 30 days or more frequently if the 5790A is operated in a
dusty environment. (Instructions for cleaning the air filter are in Chapter 8.)

2-5. Placement and Rack Mounting

You can place the 5790A on top of a workbench or mount it in a standard 19-inch wide,
24-inch (61-cm) deep equipment rack. For bench-top use, the 5790A is equipped with
non-slipping, non-marring feet. To mount the 5790A in an equipment rack, order the
accessory Y-5737 Rack Mount Kit. The rack mount kit comes with rack ears, 24-inch
slides, fasteners, and an instruction sheet.

2-6. Accessing the Fuse

To prevent instrument damage, verify that the correct fuse is
installed in accordance with the line voltage setting.
The line power fuse is accessible on the rear panel. The fuse rating label to the right of
the fuse holder (labeled F1) shows the correct replacement fuse rating for each line
voltage setting. To access the fuse, refer to Figure 2-2 and proceed as follows:
1. Disconnect line power.
2. Insert the blade of a standard screwdriver in the slot of the fuse holder labeled F1.
3. Turn the screwdriver counterclockwise until the cap and fuse pop free.
CHASSIS
GROUND
S2
S3
S4
W Caution
JOHN FLUKE MFG.CO., INC
MADE IN U.S.A
VARIOUS ASPECTS OF THIS INSTRUMENT ARE PROTECTED BY
ONE OR MORE OF THE FOLLOWING U.S. PATENTS.
4803457, 4860289, 4901052, 4716398, 4847507
NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO QUALIFIED
SERVICE PERSONNEL
TO CLEAN FILTER, REMOVE
FROM INSTRUMENT AND FLUSH
WITH WARM SOAPY WATER
CAUTION:
FOR FIRE P ROTECTION REPLACE
ONLY WITH A 250V FUSE OF INDICATED RATING.
VOLTAGE
VOLTAGE
FUSE-F1
SELECTION
SELECTION
S2 S3 S4
S2 S3 S4
100V
200V
T 1.5A
110V
220V
250V
115V
(SB)
230V
120V
240V
RS 232C
FUSE -F1
Figure 2-2. Accessing the Fuse
WARNING
GROUNDING CONNECTOR
IN POWER CORD MUST BE
CONNECTED TO ENSURE PROTECTION
FROM ELECTRIC SHOCK.
CALIBRATION MODE
FUSE-F1
SERVICE
PERIODIC
T .75A
250V
CALIBRATION STORE
(SB)
ENABLE
NORMAL
IEEE-488
Installation

Placement and Rack Mounting

add3f.eps
2
2-5