Fluke 5790A Operator's Manual page 81

Ac measurement standard
Hide thumbs Also See for 5790A:
Table of Contents

Advertisement

Press the "Remote Port Setup" softkey. Press the "Remote Port" softkey so that
"SERIAL" appears as shown below:
Press the "RS-232 Port Setup" softkey. The RS-232 port setup menu appears:
Press the softkeys to select the desired baud rate, number of data bits, number of stop
bits, and parity checking method. Press the "NEXT Menu" softkey. The second RS-232
parameter menu appears:
Press the softkeys to select the desired stall method, terminal/computer responses, and
EOL characters. Choose TERMINAL mode for interactive remote control. Responses
from the 5790A are longer and are intended for a human operator using a terminal.
Choose COMPUTER mode to get compact responses intended for use by a computer
program. It is these OMPUTER mode responses that are listed in the remote command
information and reference guide.
Use a null-modem serial interface cable such as Fluke accessory Y1702 to interconnect
your computer or terminal and the RS-232-C connector on the 5790A.
Using the Rs-232 Interface for Remote Control
5
Remote Operation
add41f.eps
add42f.eps
add43f.eps
add44f.eps
5-5