en
Troubleshooting
Warning code
Cause
255
The temperature is too low.
256
Calibration interval for DrägerSensor XXS
measuring channel 1 expired.
275
Compensation channel calibration interval
276
Calibration necessary because of over-gassing of
the compensation channel.
351
DrägerSensor XXS measuring channel 2 warming
up (warm-up phase 1).
352
DrägerSensor XXS measuring channel 2 warming
up (warm-up phase 2).
353
Measuring channel 2 concentration has drifted into
the negative range.
354
The temperature is too high.
355
The temperature is too low.
356
Calibration interval for DrägerSensor XXS
measuring channel 2 expired.
12
Remedy
Operate the gas detector within the permissible
temperature range.
Carry out a span calibration for DrägerSensor XXS
measuring channel 1, see Section 5.3 on Page 9
Carry out a calibration for DrägerSensor XXS
measuring channel 1, see Section 5 on Page 8.
Carry out a calibration for DrägerSensor XXS
measuring channel 1, see Section 5 on Page 8.
Wait until warm-up time is complete.
Wait until warm-up time is complete.
Carry out a fresh air calibration for DrägerSensor XXS
measuring channel 2, see Section 5.2 on Page 8.
Operate the gas detector within the permissible
temperature range.
Operate the gas detector within the permissible
temperature range.
Carry out a span calibration for DrägerSensor XXS
measuring channel 2, see Section 5.3 on Page 9