Setting Ramp Units; Setting Dwell Units - Barnstead F46110CM Operation Manual And Parts List

High temperature muffle furnace
Hide thumbs Also See for F46110CM:
Table of Contents

Advertisement

Note
The value set in this parameter is
always for the entire program.
To set the holdback type:
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"Hb."
3.
Press the UP or DOWN button to toggle between
"bAnd, Hi, Lo and OFF."
Holdback Value
To set the holdback value:
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"Hb.U."
3.
Press the UP or DOWN button to enter a hold-
back value.

Setting Ramp Units

Ramp units are time units which are used in "rmP.r" seg-
ments (ramp to a setpoint at degrees per second, minute
or hour) and "rmP.t" segments (ramp to setpoint in a specif-
ic amount of time). See "Setting the Segment Type" for an
explanation on hoe to set a ramp segment.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"rmP.U."
3.
Press the UP or DOWN button to toggle between
seconds, minutes and hours.

Setting Dwell Units

Dwell units are time units which are used in "dwELL" seg-
ments (amount of time to remain at a specific temperature
). See "Setting the Segment Type" for an explanation on
how to set a dwell segment.
1.
Press the PAGE button until you reach the pro-
gram list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"dwL.U."
3.
Press the UP or DOWN button to toggle
between seconds, minutes and hours.
O
PERATION
23