Setting The Target Setpoint - Barnstead F46110CM Operation Manual And Parts List

High temperature muffle furnace
Hide thumbs Also See for F46110CM:
Table of Contents

Advertisement

Caution
Minimize operation of furnace under
800°C. Element life is reduced when
operating below 800°C because the
protective layer of silica glass takes
longer to form.
4.
Press the SCROLL button until display reads,
"PrG.n."
5.
Press the UP or DOWN button to select a
program number to be linked.
6.
Press the SCROLL button until display reads,
"CYC.n."
7.
Press the UP or DOWN button to select the
number of cycles the linked program is to be
run.
End
To end or repeat a program:
1.
Press the PAGE button until you reach the
program list (ProG LiSt).
2.
Press the SCROLL button until display reads,
"tYPE."
3.
Press the UP or DOWN button until display
reads, "End."
4.
Press the SCROLL button until display reads,
"End.t."
5.
Press the UP or DOWN button to toggle
between "dwEll" (an indefinite dwell), "S OP"
(End Segment Output Power) and "rSET"
(reset).

Setting the Target Setpoint

1.
Press the PAGE button until you reach the
program list (ProG LiSt).
2.
Press the SCROLL button until display
reads, "tGt."
3.
Press the UP or DOWN button to set the
target setpoint temperature.
O
PERATION
27

Advertisement

Table of Contents
loading

Table of Contents