O
PERATION
18
FASt: Fast run through program (no/YES). The controller will
default to "no."
out.n: Even output states (OFF/ON), only used on multi-pro-
gram models.
SYnc: Synchronization (no/YES), only used on multi-program
models.
SEG.d: Flash active segment type in the lower display of the
home display (no/YES). The controller will default to "no."
ProG LiSt (Program Edit List)
PrG.n: Press the UP or DOWN ARROW to select the pro-
gram number (program number will be displayed in lower dis-
play). Only used on multi-program models.
Hb: Press the UP or DOWN ARROW to select the holdback
type [OFF (disables holdback), Lo (deviation low holdback),
Hi (deviation high holdback) or bAnd (deviation band hold-
back)] for the entire program. The controller will default to
"OFF."
Hb.U: Press the UP or DOWN ARROW to select the hold-
back value (in display units).
rmP.U: Press the UP or DOWN ARROW to toggle between
ramp units (SEc, min or Hour). Controller will default to "SEc."
dwL.U: Press the UP or DOWN ARROW to toggle between
dwell units (SEc, min or Hour). Controller will default to "SEc."
Cyc.n: Press the UP or DOWN ARROW to set the number of
program cycles (1 to 999 or cont). The controller will default to
"cont."
SEG.n: Press the UP or DOWN ARROW to select the seg-
ment number (1 -16).
tYPE: Press the UP or DOWN ARROW to select the segment
type [End (end of program), rmP.r = ramp rate (ramp to a
specified setpoint at a set rate), rmp.t = ramp time (ramp to a
specified temperature in a set time), dwEll (to maintain a con-
stant temperature for a set time), StEP (climb instantaneously
from current to specified temperature), cALL (to call a pro-
gram as a subroutine, available only on multi-program mod-
els)]. The controller will default to "End." Other parameters