Setting the standard pressure which measurement and display
8.2.4.5
values refer to for flow rate
rEF.P
select and set the required standard pressure. Setting
range: 950 to 1050 hPa in 1 hPa increments.
Setting the standard temperature which measurement and display
8.2.4.6
values refer to for flow rate
rEF.T
select and set the required standard temperature. Setting
range: 0 to 25 °C in 1 °C increments.
Setting the low flow cut-off
8.2.4.7
LFC
select and set the limit value.
Setting range: 0.1 to 0.8 Nm³/h in 0.1 Nm³/h increments.
Service functions
8.2.5
Reading min./max. values for flow rate
8.2.5.1
HI
LO
Set
HI
LO
Set
or
select; press
briefly.
HI = maximum value; LO = minimum value
Deleting memory
or
select.
press and hold it down until [----] is displayed.
Press Mode/Enter briefly.
It is a good idea to clear the memory as soon as the
instrument is used for the first time under normal working
conditions.