MS60/TS60, Check & Adjust
Step
Description
6.
Adjustment Status
Number of measurements: Shows the number of runs completed. One run
consists of a measurement in face I and face II.
a T-axis quality (1 σ): shows the standard deviation of the determined tilting
axis error. The standard deviation can be calculated from the second run
onwards.
Measure at least two runs.
7.
Next to continue with the check & adjust procedure.
8.
Select Add another calibration loop if more runs have to be added. Next
and continue with step 3.
OR
Select Finish the calibration & store the results to finish the calibration
process. No more runs can be added later. Next to view the adjustment
results.
9.
Select Finish to accept the results. No more runs can be added later.
OR
Select Redo to decline all measurements and to repeat all calibration runs.
IF the results are
to be stored
to be determined again
THEN
Next overwrites the old tilting axis error with the new one.
Redo rejects the new determined tilting axis error and
repeats the whole procedure. Refer to paragraph "Tilting
Axis Adjustment (a)".