Page 1
User Instructions Ultrasonic homogenisers Ultrasonic Homogeniser HD 4100 Valid for: HD 4050 Volume: 0.5 ml - 100 ml HD 4100 Volume: 2 ml - 200 ml HD 4200 Volume: 5 ml - 1000 ml HD 4400 Volume: 100 ml - 3000 ml 51376g GB/2018-04...
Page 2
Copyright & limit of liability This document may not be reproduced, either in full or in extracts, without prior approval of BANDELIN electronic GmbH & Co. KG, hereinafter referred to as BANDELIN. The German original is the binding version of this document. Any differences in the translation are not binding and have no legal effect.
Page 3
Identifi es instructions that must be followed in the described Handling instructions sequence. 51376g GB/2018-04 3 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Selecting a suitable probe .................... 26 4.3.1 Mounting and dismounting of the probes ..............27 4.3.1.1 Mounting and dismounting the titanium plates ............. 28 4 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 5
Taking the unit out of service ..................41 Key words ........................41 Informative appendices Spanner for mounting/dismounting Decontamination - sample copy 51376g GB/2018-04 5 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
A declaration of conformity can be requested from the manufacturer by providing the serial number. 51376g GB/2018-04 7 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Protected against access to hazardous parts Not protected with tools; protected against solid objects with a diameter of 2.5 mm or larger 51376g GB/2018-04 9 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Remote control connector • Status confi rmation External data protocol • Status inquiry RS 232 interface • Full device control 10 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 11
The inputs and outputs of the remote control connector and RS-232 interface may only be connected with devices of protection class I that are connected to the same mains supply circuit. 51376g GB/2018-04 11 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• Only use the prescribed tool for mounting and dismounting (see appendix A). • Do not use any bent probes ( unstable operation, loss of power). 12 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 13
• Repairs must only be conducted by authorised skilled personnel or by the manufacturer. • Defective parts must only be replaced with original parts or parts of the same quality! 51376g GB/2018-04 13 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Connect to a grounded socket only if type is suitable. 14 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
If necessary, an existing UW 2070 or UW 3100 can also be connected using an adapter. The type designation is then also HD 4100. 16 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
0.000 kJ Energy 0.000 kJ START STOP back to stand-by mode proceed to next set value or operating parameter 51376g GB/2018-04 17 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The pulse period duration is the sum of the turn-on and turn-off time t = t 18 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Pulsation 4.0s 4.0s Temperature 31°C No sensor: ---°C Energy 0.000 kJ START STOP back to stand-by mode 51376g GB/2018-04 19 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Note: In this operating mode no more changes of settings are possible within the stand-by mode or ultrasonic operation. 22 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• The displayed value for energy is calculated on the basis of HF energy applied and, after taking into the account the effi ciency value, can be used as an equivalent for the acoustic ultrasound energy. 51376g GB/2018-04 23 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Set the immersion depth for the probe (approx. 10 -20 mm). If needed, connect the temperature sensor and position it in the medium. 24 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 25
Clean the probe and check the wear on the sound emitting surface at regular intervals, see chapter 5.1. 51376g GB/2018-04 25 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
After technical updating of the probes, the data may be changed. The stored data in the program are always valid! 26 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The mounting surfaces are the contact surfaces between the individual accessory parts, e.g. between the standard horn and the probe. 51376g GB/2018-04 27 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
One sickle spanner is part of the s cope of delivery of one standard horn. Dismounting Ultrasonic transducer Standard horn Vigorously turn sickle spanner HS 40/42 long, in the opposite direction 30 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 31
The ultrasonic oscillating system and its electrical connections would be damaged as a result. 51376g GB/2018-04 31 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• The exterior of the housing should be wiped down with a damp cloth and left to dry or wiped dry. 32 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
During extended periods without use, the homogeniser should be stored in a cool, dry place. The ultrasonic transducer should be covered in order to protect the electronics from outside contamination. 51376g GB/2018-04 33 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 35
The ultrasonic oscillating system and its electrical connections would be damaged as a result. 51376g GB/2018-04 35 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Critical faults are recognised by the device and signalled by a red LED and an intermittent signal tone (3 times), and displayed with an error number, see next page. 36 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 37
51376g GB/2018-04 37 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 38
If it is not possible to rectify the fault using these short instructions, please contact your local specialist dealer or write us to the address below. 38 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
03.01 - 2016-08-085 The software version provided here (bottom right) is only an example, the actual specifi cations may differ. 51376g GB/2018-04 39 / 48 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
SH 200 G SH 200 G SH 400 G SH 70 G Connection only possible using an adapter. The variable amplitudes are comparable to the HD 2070. 40 / 48 51376g GB/2018-04 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 43
7, 12, 28, 32, 33, 37, 38 Tool 12, 27, 38 Torque 28, 29 Type plate 6, 9, 14 BANDELIN electronic GmbH & Co. KG www.bandelin.com Zertifi ziert nach Heinrichstraße 3 – 4 info@bandelin.com EN ISO 9001:2008 12207 Berlin +49 30 768 80-0...
Page 45
Spanner for mounting/dismounting Probes and standard horns are highly-sensitive parts that must be mounted and dismounted with the greatest amount of care. For this reason, only use the following spanners for the mounting/dismounting of probes from standard horns, as well as of standard horns from the ultrasonic transducer: Spanner Type To mount/dismount probes ...
Page 46
Decontamination - sample copy Certifi cate of Decontamination ! ! ! CAUTION! ! ! This form must be visibly affi xed to the outside of the package! This "Certifi cate of Decontamination" is intended to protect the occupational health and safety of our employees pursuant to the German Protection against Infection Act and trade association accident prevention regulations.
Page 47
Certifi cate of Decontamination ! ! ! CAUTION! ! ! This form must be visibly affi xed to the outside of the package! Legally binding statement I/We hereby declare that the device and accessories found in this package have been cleaned and/or disinfected pursuant to current laws and regulations and that the information provided in this declaration is correct and complete: Company / Institution: Street and number:...
Page 48
Important: The Instructions for Use in this and other languages, as well as further information, can be found on the enclosed CD.
Need help?
Do you have a question about the Sonopuls HD 4050 and is the answer not in the manual?
Questions and answers