Page 1
User instructions Ultrasonic homogenisers Ultrasonic homogeniser HD 2070.2 valid for: HD 2070.2 Volumes: 1 ml - 200 ml HD 2200.2 Volumes: 2 ml - 1000 ml 51506-002 GB/2020-10...
Page 2
Copyright and disclaimer This document may not be reproduced, whether in full or in part, without the prior approval of BANDELIN electronic GmbH & Co. KG, hereinafter “BANDELIN”. The German-language original is the binding version of this document. Any diff erences in the translation are not binding and have no legal eff...
Page 3
Identifi es instructions that must be followed Instructions in the described sequence. 51506-002 GB/2020-10 3 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Mounting and dismounting of micro and tapered tips and long probes ......26 Dismounting and mounting of the standard and booster horns ........27 4 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 5
Optional accessories ....................36 Consumable materials ....................37 Decommissioning ......................37 Keyword index ......................38 General information Tools for mounting/dismounting 51506-002 GB/2020-10 5 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
A declaration of conformity can be requested from the manufacturer by providing the serial number. 51506-002 GB/2020-10 7 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
In the case of highly-viscous media, this can also result in the desired amplitude not always being reached. 8 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 9
Technical measurement checks, MTK (section 11): not applicable e-procurement e-cl@ss: HD device classifi cation: 32-08-02-02 HD accessories classifi cation: 32-08-02-05 51506-002 GB/2020-10 9 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
L Ultrasonic operation OFF 5V H H Ultrasonic operation ON Output As with contact 4, but with signal inverted 10 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
It is suitable for use in institutions and other areas that are directly connected to a public low-voltage power supply, e.g., medical laboratory facilities. 51506-002 GB/2020-10 11 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• Before any mounting or dismounting (Chapter 4.4) of probes or standard/booster horns as well as accessories, turn off the device and disconnect the ultrasonic transducer from the ultrasonic generator. 12 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 13
• Repairs are only to be conducted by authorised skilled personnel or by the manufacturer. • Defective parts must only be replaced with original parts or parts of the same quality! 51506-002 GB/2020-10 13 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• Avoid any contact between the contact pins on the ultrasonic transducer's plug, since the piezo elements could generate electric loads through temperature fl uctuations with a possibility of slight electric shocks. 14 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
(inverse view) of the selected parameter, the desired value is set by turning the rotary knob. By pressing the “START/STOP” button, the editing mode is exited. 16 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
1 to 99 minutes 1 to 59 seconds proceed to next set value or operating parameter START back to stand-by mode STOP 51506-002 GB/2020-10 17 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
STOP The pulse period duration is the sum of the turn-on and turn-off time t = t 18 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
1 and 255, and can be set by turning the rotary knob. Smaller values result in a darker image and larger values result in a lighter image. The default value is 240. 51506-002 GB/2020-10 19 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Once the maximum displayable time has been reached and exceeded, the time display will start again at 00:00. 20 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Position the sonication vessel below the ultrasonic transducer in such a manner that the probe Set the immersion depth for the probe (approx. 1 - 2 cm). 51506-002 GB/2020-10 21 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 22
Clean the probe and check the wear on the sound emitting surface at regular intervals, see Chapter 5.1. 22 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
100 % amplitude setting Amplitude values may diff er slightly from the indicated value due to material and manufacturing tolerances. 51506-002 GB/2020-10 23 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The mounting surfaces are the contact surfaces between the individual accessory parts, e.g. between the standard or booster horn and the probe. 24 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
fi rmly in the opposite direction. Ultrasonic transducer Standard and/or booster horn Vigorously turn sickle spanner HS 40/42 long, in the opposite direction 51506-002 GB/2020-10 27 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 28
The ultrasonic oscillating system and its electrical connections would be damaged as a result. 28 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• Housing is to be wiped off only from the outside, use a suitable surface disinfectant if needed. Afterwards, allow to dry off or wipe dry. 51506-002 GB/2020-10 29 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
During extended periods without use, the homogeniser should be stored in a cool, dry place. The ultrasonic transducer should be covered in order to protect the electronics from outside contamination. 30 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
>2s start scan frequency scan frequency START contrast contrast STOP V xxxx 199547 Hz V xxxx 199547 Hz 51506-002 GB/2020-10 31 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 32
The ultrasonic oscillating system and its electrical connections would be damaged as a result. 32 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• If it was possible to remedy the error, the display indicator should be deleted by turning off and on again. 51506-002 GB/2020-10 33 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 34
If it is not possible to rectify the error using these short instructions, please contact your local specialist dealer or write us to the address below. 34 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The form can be downloaded from our website as a PDF fi le: www.bandelin.com – Download ... 51506-002 GB/2020-10 35 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
The sonication of liquids generates noise. The sound proof box mutes the sound emission up to approximately 30 dB(AU) LS 40 36 / 42 51506-002 GB/2020-10 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
• Metal accessories such as probes or standard horns should be decontaminated and disposed of as metal waste. • The packing is recyclable. 51506-002 GB/2020-10 37 / 42 1 electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • info@bandelin.com...
Page 40
21, 22, 24, 27, 28, 30, 31, 32, 33, 34, 3, 13, 33 Viscosity 8, 20, 21 Voltage source 11, 12 BANDELIN electronic GmbH & Co. KG www.bandelin.com Zertifi ziert nach Heinrichstraße 3 – 4 info@bandelin.com ISO 9001 12207 Berlin ...
Page 41
Tools for mounting/dismounting Probes, standard horns and booster horns are highly-sensitive parts that must be mounted and dismounted with the greatest care. For this reason, only use the following keys for mounting/dismounting of probes from standard or booster horns, as well as of standard or booster horns from the ultrasonic transducer: Spanner Usage Type...
Page 42
Please note: The user instructions in this and other languages, as well as further information, can be found on the enclosed CD.
Need help?
Do you have a question about the SONOPULS and is the answer not in the manual?
Questions and answers