Emerson Rosemount 8750W Reference Manual page 107

Magnetic flowmeter platform for utility, water, and wastewater applications
Hide thumbs Also See for Rosemount 8750W:
Table of Contents

Advertisement

Reference Manual
00809-0300-4750, Rev CA
DI/O 1 control
LOI menu path
Traditional Fast Keys
Device dashboard
This parameter configures the auxiliary output channel 1. It controls whether channel 1 will
be a discrete input or discrete output on terminals 5(-) and 6(+). Note that the transmitter
must have been ordered with the auxiliary output suite (option code AX) to have access to
this functionality.
Discrete input 1
LOI menu path
Traditional Fast Keys
Device dashboard
This parameter displays the configuration for channel 1 when used as a discrete input. Refer
to the list above for available discrete input functions.
Discrete output 1
LOI menu path
Traditional Fast Keys
Device dashboard
This parameter displays the configuration for channel 1 when used as a discrete output.
Refer to the list above for available discrete output functions.
Channel 2
Channel 2 is only available as discrete output only.
Discrete output 2
LOI menu path
Traditional Fast Keys
Device dashboard
This parameter displays the configuration for channel 2. Refer to the list above for available
discrete output functions.
Advanced Configuration Functionality
Advanced Configuration Functionality
Field Mount: Detailed Setup, Output Config, DI/DO Config, DI/O 1,
DI/O Control
Wall Mount: AUX. FUNCTION
1, 4, 2, 3, 1, 1
2, 2, 4, 1, 1
Field Mount: Detailed Setup, Output Config, DI/DO Config, DI/O 1, DI 1
Wall Mount: AUX. FUNCTION
1, 4, 2, 3, 1, 1, 3
2, 2, 4, 1, 2
Field Mount: Detailed Setup, Output Config, DI/DO Config, DI/O 1, DO 1
Wall Mount: AUX. FUNCTION
1, 4, 2, 3, 1, 2, 4
2, 2, 4, 1, 3
Field Mount: Detailed Setup, Output Config, DI/DO Config, DO 2
Wall Mount: AUX. FUNCTION
1, 4, 2, 3, 2
2, 2, 4, 2
March 2016
95