S/Impedances Menu; Glass; Reference; Test - Hach POLYMETRON 9135 Operator's Manual

Ph/redox measuremen
Table of Contents

Advertisement

Transmitter 9135 - pH/redox measurement
IMPEDANCES

GLASS

REFERENCE

FREQ. : 24.0 h

TEST

Select
GLASS
YES
REF. : OUI
LOWER : XXXX
UPPER :XXXX
Select
REFERENCE
YES
REF. : OUI
LOWER : XXXX
UPPER :XXXX
Select
TEST
GLASS
XXX.XX
REFERENCE
XX.XX
Page 54

S/IMPEDANCES Menu

Define the lower and upper limits of the electrode
impedances for the default detection. They should
be defined experimentally .
FREQ.
: permits to define the impedance
measurement frequency (limits :
0,1 H to 24 H).
GLASS
Permits an impedance measurement.
The glass electrode impedance measurement is
compensated at 25°C.
REFERENCE
Permits an impedance measurement.
LOWER
: defines the lower limit for the
impedance measurement. This limit
is used for the alarm system and
the bargraph DISP4.
UPPER
: defines the upper limit for the
impedance measurement. This limit
is used for the alarm system and
the bargraph DISP4.
TEST
Visualizes the impedance measurement value.