USING THE OPTIONAL WRIST RESTS

There are two optional wrist rests avail-
able; standard  and temporal .
To use the wrist rest, insert the support
tube (A) into the socket in the Fowler head
piece assembly. Turn knob (B) clockwise
to secure the wrist rest assembly.
To adjust the height of the wrist rest, turn
knob (C) counterclockwise to loosen it.
Raise or lower the wrist rest to the desired
height, and turn the knob clockwise to
tighten it and hold the wrist rest in place.
The "U" shaped rest (D) can be pivoted up
and away from the patient when the wrist
rest is not in use.

USING THE OPTIONAL TETHERED I.V. POLE

To use the tethered I.V. pole:
1. Remove the I.V. pole from the storage trough under the
litter and insert into the receptacle on the corner of the
litter frame.
2. To raise the height of the pole, turn knob (A) count-
erclockwise and pull up on the telescoping portion (B)
of the pole to raise it to the desired height.
3. Turn knob (A) clockwise to lock the telescoping portion
in place.