Complete the parameter entry by pressing LEFT
measured value.
3.3.2: Simulate DO (F = End)
Simulation of the digital output signal
Value
Select the applied output value (relay on or relay off).
Complete the parameter entry by pressing LEFT
alarm/limit.
3.3.3: Error Simulation for Micro Controller High Word
For factory service purposes only
3.3.4: Error Simulation for Micro Controller Low Word
For factory service purposes only
3.3.5: Error Simulation for DSP Controller High Word
For factory service purposes only
3.3.6: Error Simulation for DSP Controller Low Word
For factory service purposes only
3.4: Sensor Variables
You can read out internal device data with this parameter group. The displayed values
depend on the respective application. You can access the following data:
3.4.1: Raw Value (for service purposes only)
Distance from the flange to material
The measured distance from the flange to the surface of the material.
3.4.2: Echo Amplitude
Measure of quality of reflection
This dimensionless value is an absolute measure of the strength of reflection at the
material. Its display can be evaluated as follows:
x > 1:
1 > x > 0.5:
0.5 > x > 0.05: satisfactory
x < 0.05:
7ML19985JC02
Relay on
Relay off
End
very good
good
uncertain
SITRANS LR 400 (Solids) – INSTRUCTION MANUAL
so the analog output again gives the
so the digital output again gives an