Electromagnetic Emissions

Guidance and manufacturer's declaration; electromagnetic emissions
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
Emissions
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
Not applicable
emissions
IEC 61000-3-2
Voltage
Not applicable
fluctuations/
flicker emissions
IEC 61000-3-3

Electromagnetic Immunity

Guidance and manufacturer's declaration; electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the
product should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
Surge
±1 kV line(s) and
neutral
IEC 61000-4-5
Voltage dips, short
<5% U
interruptions and
(>95% dip in U
voltage variations
for 0,5 cycle
on power supply
40% U
input lines
(60% dip in U
IEC 61000-4-11
for 5 cycles
70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level
T
16 | Electromagnetic Compatibility Tables
Electromagnetic environment-- guidance
The product uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The product is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Compliance
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV line(s) and
neutral
<5% U
T
T
)
(>95% dip in U
T
for 0,5 cycle
40% U
T
T
)
(60% dip in U
)
T
T
for 5 cycles
70% U
T
T
)
(30% dip in U
)
T
T
for 25 cycles)
<5% U
T
T
)
(>95% dip in U
T
for 5 sec
Not applicable
Electromagnetic environment - guidance
The product uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The product is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Mains power quality should be that of a typical commercial or
hospital environment
Mains power quality should be that of a typical commercial or
)
hospital environment. If a dips or an interruption of mains
T
power occurs, the current of the product may be dropped off
from normal level, it may be necessary to use uninterruptible
power supply or a battery.
)