Power Connector and Pinout
24 VDC Connector
Electrical Symbols
Equipotentiality
This symbol appears next to the display Potential Equalization Conductor (ground post).
Closed (On) Switch
This symbol appears below the closed, or on, side of the display On/Off switch.
Open (Off) Switch
This symbol appears below the open, or off, side of the display On/Off switch.
Optional 5 VDC Fiber Power Cable
J1 Pinout
2
1
J1
To purchase the optional 5 VDC Fiber Power Cable (P/N 35X0100) contact:
North America and Asia Pacific: customercare@ndssi.com
Cable Bend Radius
We recommend that the bend radius of metallic cables be no less than 63 mm (2.5 inches) or 7 times
the diameter of the cable whichever is greater. The bend radius of Fiber Optic cables should be no less
than 10 times the diameter of the cable. Sharper bends can damage the cable, and/or degrade the
video signal.
3
1
2
Pin Name
Description
1
+24 VDC
Power Input
2
GND
Ground
3
Shield
Ground
5 VDC OUT
NDS 35X0100
Connector Center (1)
J1
+5 VDC
J2
+5 VDC
Europe: customercare-emea@ndssi.com
5 VDC FIBER
Shield (2)
Return
Return
Connector Panel Overview | 9
J2 Pinout
1
2
J2