Warning: To avoid risk of electric shock, this equipment must be only connected to a supply mains with protective earth. The installation shall only be carried out by NDS authorized and trained personnel. This device is only intended to be used in professional healthcare environment.
Power Supply For Customers in North America EndoVue series monitors comply to the above standards only when used with the supplied hospital grade power supply. Note: This power supply is a part of the medical device. Model BridgePower BM060S24F AC Input 100 - 240 Volts, 50 to 60 Hz DC Output 24 Volts, 2.7 Amps For Customers in China This power supply is to be used with this monitor.
Safety Precautions External equipment intended for connection to signal input/output or other connectors shall meet the level of safety equivalent to ME equipment complying with the relevant standard. In addition, all such combinations- systems shall comply with the standard. Equipment not complying with UL/EN/IEC 60601-1 shall be kept outside the patient environment, as defined in the standard.
NDS neither assumes nor authorizes any person to assume for it any other liabilities in conjunction with and/or related to the sale and/or use of its products. To ensure proper use, handling and care of NDS products, customers should consult the product specific literature, instruction manual, and/or labeling included with the product or otherwise available.
Subject to the conditions set forth below, NDS agrees to repair or replace any part of the enclosed unit for the length of period indicated on the chart below:...
Page 12
NDS's liability is limited to the cost of repair or replacement of the product. NDS shall not be liable for: 1. Damage to other property caused by any defects in the product, damages based upon inconvenience, loss of use of the product, loss of time, loss of profits, loss of business opportunity, loss of goodwill, interference with business relationships, or other commercial loss, even if advised of the possibility of such damages.
VESA Compliant Display Mounting EndoVue series displays are compliant with the VESA Mounting Interface Standard, with MIS hole mount patterns configured to size and weight of the display, suitable for stand, wall, or armature mounting. MIS-D 100 x 100 mm and MIS-E 200 x 100 mm interfaces are shown below. M4 X 0.7 100 mm Thread...
Enclosure Assembly and Cleaning Cable Cover Installation 1. Connect power, control, and video cables before installing the cable cover. 2. Align the cable cover to the cable well recess on the back of the display. 3. Slide the cable cover forward into the recess, with cables positioned under the cut-out. 4.
Display User Interface Display Keypad The Display Keypad is centered on the lower front surface of the display enclosure, providing controls for adjustment of display parameters using the On Screen Display (OSD) Menu system. INPUT MENU SCROLL Menu Navigation INPUT and MENU Buttons: Open OSD Menus To open the Input Menu, press the INPUT button.
SmartSync™/Alternative Modes (VGA/RGBS, YPbPr only) NDS proprietary SmartSync technology examines incoming signals to automatically display the video image in its proper format. Alternative Modes are used to manually distinguish between modes whose timing characteristics are very close.
Video Format (Composite, SDI, S-Video only) Auto: Examines incoming signals to automatically display the video image in NTSC or PAL format. NTSC/PAL: Press the buttons to manually select. Color Menu Gamma To select a gamma setting, press the buttons.. 1.8, 2.0, 2.2, 2.4 or 2.6: Preset gamma values Video: Linear gamma Look Up Table (LUT) PACS: DICOM gamma LUT...
Operating Hours Displays in hours and minutes the accumulated time the monitor has been in use. Version number of the display firmware. Defaults Menu Factory Defaults To return all changed parameter settings to factory preset values, press the button. The “Restoring Factory Defaults” message displays while processing.
Intensity (Chrominance) Composite Pin Name Description DATA GROUND Sleeve Firmware Connector and Pinout Mini USB Flash Upgrade Cable, part number 35Z0047, is available from NDS. Pin Name Description +5 VDC D Minus Data Return D Plus Data Transmit Ground 8 | Connector Panel Overview...
This symbol appears below the open, or off, side of the display On/Off switch. Optional 5 VDC Fiber Power Cable J1 Pinout J2 Pinout 5 VDC OUT NDS 35X0100 5 VDC FIBER Connector Center (1) Shield (2) +5 VDC Return...
Specifications Specifications Specifications are subject to change without notice. Contact factory for current specifications. EndoVue 19" EndoVue 21" EndoVue 24" EndoVue 32" 19.0 (483 mm) 21.5 in. (546 mm) 24.1 in. (612 mm) 31.5 in. (800 mm) Viewing Area (Diagonal) Typical Luminance (cd/m 1280 x 1024 1920 x 1080...
Troubleshooting Image Size is Large for the Screen (VGA, RGBS, YPbPr analog inputs only) If the image does not appear to be the correct format, then SmartSync must be run. To run SmartSync, press the MENU button. In the Picture Menu, press the SCROLL button to highlight SmartSync and press the button.
Electromagnetic Compatibility Tables All medical electronic devices must conform to the requirements of IEC 60601-1-2. Precautions, adherences to the Electromagnetic Compatibility (EMC) guideline information provided in this manual and verification of all medical devices in simultaneous operation are required to ensure the electromagnetic compatibility and co- existence of all other medical devices prior to a surgical procedure.
Electromagnetic Emissions Guidance and manufacturer’s declaration; electromagnetic emissions The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Emissions Compliance Electromagnetic environment-- guidance RF emissions Group 1...
Guidance and manufacturer’s declaration; electromagnetic immunity (continued) The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Portable and mobile RF communications equipment should be used no closer to any part of the product, including cables, than the recommended separation distance calculated from the...
+81 3 5753 2466 info@ndssi.jp Oriental Media Center Suite 2302, Tower C No.4, Guang Hua Road Chao Yang District Beijing, 100026 China +86 10 8559 7859 info@ndssi.com.cn NDS Quality System ISO 9001:2008 and ISO 13485:2003 FDA Registration #2954921 60A0585 Rev C...